TY - JOUR
T1 - The Silent Acoustic Noise Protocol of Magnetic Resonance Imaging Examination in the Case of Head Image
AU - Muzamil, Akhmad
AU - Nurdin, Dezy Zahrotul Istiqomah
AU - Rulaningtyas, Riries
AU - Astuti, Suryani Dyah
N1 - Publisher Copyright:
© (2023), (Mashhad University of Medical Sciences). All Rights Reserved.
PY - 2023
Y1 - 2023
N2 - Introduction: Health examinations are performed every six months. X-rays, magnetic resonance imaging (MRI), ECGs, and blood tests are all part of health examinations. In this investigation, the silent T2 Fast-Spin-Echo (FSE) and Gradient-Recalled Echo (GRE) MRI head examination sequences are compared. Noise is produced during an MRI test in addition to images. Material and Methods: This research was conducted by adjusting the parameters on the MRI, such as time repeat (TR), time echo (TE), and echo train length (ETL). Then, the resulting silent sequence image is simulated with a simulation program. Results: The variation of TR 440 with TE 24 in the GRE sequence for the white matter (WM) tissue has the highest signal to noise ratio (SNR) value. The cerebispinal fluid (CSF) tissue is also in the TR 560/TE 20 variant at the same time. Then, variations of TR 3360, TE 97, and ETL 33.6 have the highest peak signal to noise ratio (PSNR) measurement results in the WM or CSF tissue. Conclusion: According to the study's findings, the average sound intensity level and mean square error (MSE) value produced by the GRE sequence protocol are less than those produced by the T2 FSE sequence protocol. While this is the case, the GRE sequence protocol generates an average PSNR value that is higher than the FSE T2 sequence protocol. The T2 FSE sequence with variations of TR 3360, TE 97, and ETL 33.6 may then be observed to be the best with the ideal noise level and SNR value.
AB - Introduction: Health examinations are performed every six months. X-rays, magnetic resonance imaging (MRI), ECGs, and blood tests are all part of health examinations. In this investigation, the silent T2 Fast-Spin-Echo (FSE) and Gradient-Recalled Echo (GRE) MRI head examination sequences are compared. Noise is produced during an MRI test in addition to images. Material and Methods: This research was conducted by adjusting the parameters on the MRI, such as time repeat (TR), time echo (TE), and echo train length (ETL). Then, the resulting silent sequence image is simulated with a simulation program. Results: The variation of TR 440 with TE 24 in the GRE sequence for the white matter (WM) tissue has the highest signal to noise ratio (SNR) value. The cerebispinal fluid (CSF) tissue is also in the TR 560/TE 20 variant at the same time. Then, variations of TR 3360, TE 97, and ETL 33.6 have the highest peak signal to noise ratio (PSNR) measurement results in the WM or CSF tissue. Conclusion: According to the study's findings, the average sound intensity level and mean square error (MSE) value produced by the GRE sequence protocol are less than those produced by the T2 FSE sequence protocol. While this is the case, the GRE sequence protocol generates an average PSNR value that is higher than the FSE T2 sequence protocol. The T2 FSE sequence with variations of TR 3360, TE 97, and ETL 33.6 may then be observed to be the best with the ideal noise level and SNR value.
KW - GRE
KW - Health
KW - MRI
KW - Noise Level
KW - T2 FSE
UR - http://www.scopus.com/inward/record.url?scp=85186120041&partnerID=8YFLogxK
U2 - 10.22038/IJMP.2023.67732.2179
DO - 10.22038/IJMP.2023.67732.2179
M3 - Article
AN - SCOPUS:85186120041
SN - 2252-0309
VL - 20
SP - 351
EP - 358
JO - Iranian Journal of Medical Physics
JF - Iranian Journal of Medical Physics
IS - 6
ER -