Simulation of leakage current measurement on medical devices using helmholtz coil configuration with different current flow

Research output: Contribution to journalConference articlepeer-review

6 Citations (Scopus)

Abstract

Leakage current measurement which can follow IEC standard for medical device is one of many challenges to be answered. The IEC 60601-1 has defined that the limit for a leakage current for Medical Device can be as low as 10 μ A and as high as 500 μ A, depending on which type of contact (applied part) connected to the patient. Most people are using ELCB (Earth-leakage circuit breaker) for safety purpose as this is the most common and available safety device in market. One type of ELCB devices is RCD (Residual Current Device) and this RCD type can measure the leakage current directly. This work will show the possibility on how Helmholtz Coil Configuration can be made to be like the RCD. The possibility is explored by comparing the magnetic field formula from each device, then it proceeds with a simulation using software EJS (Easy Java Simulation). The simulation will make sure the concept of magnetic field current cancellation follows the RCD concept. Finally, the possibility of increasing the measurement's sensitivity is also analyzed. The sensitivity is needed to see the possibility on reaching the minimum leakage current limit defined by IEC, 0.01mA.

Original languageEnglish
Article number012004
JournalJournal of Physics: Conference Series
Volume853
Issue number1
DOIs
Publication statusPublished - 7 Jun 2017
EventInternational Conference on Physical Instrumentation and Advanced Materials, ICPIAM 2016 - Surabaya, Indonesia
Duration: 27 Oct 2016 → …

Fingerprint

Dive into the research topics of 'Simulation of leakage current measurement on medical devices using helmholtz coil configuration with different current flow'. Together they form a unique fingerprint.

Cite this