Dose evaluation of organs at risk (OAR) cervical cancer using dose volume histogram (DVH) on brachytherapy

R. Arif Wibowo, Bambang Haris, Dan Inganatul Islamiyah

Research output: Contribution to journalConference articlepeer-review

3 Citations (Scopus)

Abstract

Brachytherapy is one way to cure cervical cancer. It works by placing a radioactive source near the tumor. However, there are some healthy tissues or organs at risk (OAR) such as bladder and rectum which received radiation also. This study aims to evaluate the radiation dose of the bladder and rectum. There were 12 total radiation dose data of the bladder and rectum obtained from patients' brachytherapy. The dose of cervix for all patients was 6 Gy. Two-dimensional calculation of the radiation dose was based on the International Commission on Radiation Units and Measurements (ICRU) points or called DICRU while the 3-dimensional calculation derived from Dose Volume Histogram (DVH) on a volume of 2 cc (D2cc). The radiation dose of bladder and rectum from both methods were analysed using independent t test. The mean DICRU of bladder was 4.33730 Gy and its D2cc was4.78090 Gy. DICRU and D2cc bladder did not differ significantly (p = 0.144). The mean DICRU of rectum was 3.57980 Gy and 4.58670 Gy for D2cc. The mean DICRU of rectum differed significantly from D2cc of rectum (p = 0.000). The three-dimensional method radiation dose of the bladder and rectum was higher than the two-dimensional method with ratios 1.10227 for bladder and 1.28127 for rectum. The radiation dose of the bladder and rectum was still below the tolerance dose. Two-dimensional calculation of the bladder and rectum dose was lower than three-dimension which was more accurate due to its calculation at the whole volume of the organs.

Original languageEnglish
Article number012013
JournalJournal of Physics: Conference Series
Volume853
Issue number1
DOIs
Publication statusPublished - 7 Jun 2017
EventInternational Conference on Physical Instrumentation and Advanced Materials, ICPIAM 2016 - Surabaya, Indonesia
Duration: 27 Oct 2016 → …

Fingerprint

Dive into the research topics of 'Dose evaluation of organs at risk (OAR) cervical cancer using dose volume histogram (DVH) on brachytherapy'. Together they form a unique fingerprint.

Cite this